61850 • Starting Scanning Pull the rubber sheath out gently at the exposed end and screw the sealing rings tight to the rectal tube (picture 3).Fil
71850 • Starting ScanningFill the plastic hood with approximately 10-20ml of degassed (lukewarm boiled) water through the valve on the rectal tube.No
81850 • Transurethral Scanning with Type 1850 The two small fins on the right of the black grip (Fig. 9.) are designed to anchor the rectosig-moido
91850 • Transrectal Scanning with Type 1850Ensure the bladder is kept well distended to avoid the transducer touching the wall.To start scanning, pre
101850 • Puncture Facilities Puncture FacilitiesPuncture and biopsy are possible with Type 1850. The puncture attachment is illustrated in the foll
111850 • Performing Puncture and BiopsyNote: Sterile, disposable sheaths are recommended for intraoperative use; and in the U.S.A. it is recommended
121850 • 3D Scanning with Type 1850 Refer to the applicable scanner user guide for the method of changing which row of puncture markers appears on
131850 • 3D Scanning with Type 1850• The Variable Friction Arm UA0553.Fig. 13. Type 1850 and Pullback Mover UA0552 mounted on Variable Friction Arm U
141850 • 3D Scanning with Type 1850 • A variable friction wheel (D).One of the two fixation clamps (B or C) is used to secure the support arm to th
151850 • DisposalAttaching the Instrument Clamp to the Support ArmThe clamp has a hand-screw (F) and a security pin (G) on the upper jaw, which contr
WORLD HEADQUARTERSMileparken 34DK-2730 HerlevDenmarkTel.:+45 44528100 / Fax:+45 44528199 www.bkmed.comEmail: [email protected] you have comments about t
161850 • Disposal
Endosonic Probe Type 1850 Introduction 1Scanning Plane 1General Information 2Caring for the Transducer 2Cleaning and Disinfection 2Starting Scanni
Attaching the Fixation Clamp to the Support Arm 14Attaching the Instrument Clamp to the Support Arm 15Maneuvering the Support Arm 15Disposal 15
1Endosonic Probe Type 1850IntroductionThis is the user guide for Endosonic Probe Type 1850 and must be used together with Trans-ducer Care, Cleaning
21850 • General Information General InformationProduct specifications for this transducer can be found in the Product Data sheet that accompanies t
31850 • Starting ScanningStarting ScanningAll equipment must be cleaned and disinfected before use.Connecting the TransducerUnscrew the connector plu
41850 • Starting Scanning Assembling Type 1850 for Transurethral UseType 1850 can use a Storz 24Fr. resectoscope, which is shown in Fig. 3. The res
51850 • Starting Scanningthreaded collar of the rectal tube with the nodule on the probe and push firmly into position. Turn the collar on the rectal